DESCRIPTION

The V9-14 Voltage Converter Set is intended for very precise measurement of ac voltage rms values. It comprises the Unit of Thermoelectric Conversion of ac voltage and current to dc voltage, the Unit of Auxiliary Resistors and the Indicator Unit.

The Thermoelectric Conversion Unit is built on the base of multielement differential film thermal converter that provides a dc voltage signal independent of frequency and shape and is proportional to rms value of ac voltage.

The Auxiliary Resistors Units are connected to the electrothermal current converter input at high level signal voltage (³ 50 V) and allow signal current value decrease.

The Digital Indicator Unit receives and processes output voltage of connected Thermoelectric Conversion Unit and displays measurement results. Besides this unit contains a two-polarity dc voltage multivalue measure, which is used for thermoelectric conversion units, calibration and a computing device, which makes necessary measurement results corrections. The V9-14 Voltage Converters Set is supplied in three options:

– for low (0.5–30 V) and high (50–1000 V) level ac voltage measurements (Option No 1);

– for low (0.5–30 V) level ac voltage measurements (No 2);

– for high (50–100 V) level ac voltage to dc voltage conversion (No 3).

GENERAL

Power: 220±22 V, 50±0.5 Hz, mains harmonics content up to 5 %

Power consumption: not more than 30 VA

MTBF: 10.000 hours

Dimensions and Weight Measurement Unit 230х185х76 mm; 3 kg

P1-P5 and V9-14-0.8А Converters 30х40 mm, length 100 mm; 0.5 kg

BDR-50; BDR-100; BDR-300 and BDR-500 UnitsÆ 55 mm, length 55 mm; 0.5 kg

BDR-1000 Unit Æ55 mm length 110 mm; 0.5 kg

SPECIFICATIONS

Characteristics

Characteristics value

Option No 1

Option No 2

Option No 3

Voltage range

0.25–1000 V
at UR=0.5; 1; 3; 10; 30; 50;
100; 300; 1000 V subranges

0.25–30 V
at UR= 0.5; 1; 3;
10; 30 V subranges

25–1000 V at
UR=50; 100; 300;
500; 1000 V subranges

Frequency range

10 Hz–100 MHz
at 0.5; 1; 3 V subranges
10 Hz–30 MHz
at 10 and 30 V subranges;
10 Hz–1 MHz
at 50 and 100 V subranges
10 Hz–100 kHz
at 300; 500 and
1000 V subranges

10 Hz–100 MHz
at 0.5; 1; 3 V subranges
10 Hz–30 MHz
at 10 and 30 V subranges

10 Hz–1 MHz
at 50 and
100 V subranges
10 Hz–100 kHz
at 300; 500 and
1000 V subranges

Permissible basic conversion accuracy at signal dc/ac voltage conversion at calibration frequency 1 kHz

±0.01 %

±0.01 %

±0.01%

Permissible basic conversion accuracy at transition of calibration frequency input signal to normal frequency range signal 0.02–20 kHz

±0.01 %

±0.01 %

±0.01%

Permissible basic conversion accuracy at transition of calibration frequency input signal to operating frequency range signal

at 0.5 и 1 V subranges:
±0.02 % (up to 0.1 MHz)%
±0.1 % (up to 1 MHz) %
±0.2 % (up to 10 MHz) %
±0.3 % (up to 30 MHz) %
±1.5 % (up to 100 MHz) %

at 0.5 и 1 V subranges:
±0.02 % (up to 0.1 MHz)%
±0.1 % (up to 1 MHz)
±0.2 % (up to 10 MHz)%
±0.3 % (up to 30 MHz)%
±1.5 % (up to 100 MHz)%

<tr bgcolor="#dfdfdf">

at 3 V subrange:
±0.01 % (up to 0.1 MHz)%
±0.05 % (up to 1 MHz)%
±0.2 % (up to 10 MHz)%
±0.3 % (up to 30 MHz)%
±1.5 % (up to 100 MHz)%

at 3 V subrange:
±0.01 % (up to 0.1 MHz)%
±0.05 % (up to 1 MHz)%
±0.2 % (up to 10 MHz)%
±0.3 % (up to 30 MHz)%
±1.5 % (up to 100 MHz)%

at 10 и 30 V subranges:
±0.01 % (up to 0.1 MHz)%
±0.05 % (up to 1 MHz) %
±0.2 % (up to 10 MHz) %
±0.3 % (up to 30 MHz) %

at 10 и 30 V subranges:
±0.01 % (up to 0.1 MHz)%
±0.05 % (up to 1 MHz)%
±0.2 % (up to 10 MHz)%
±0.3 % (up to 30 MHz)%

at 50, 100 V subranges:
±0.01 % (up to 0.1 MHz)%
±0.3 % (up to 1 MHz) %

at 50, 100 V subranges:
±0.01 % (up to 0.1 MHz) %
±0.3 % (up to 1 MHz) %

at 300 V subrange:
±0.01 % (up to 0.05 MHz)%
±0.02 % (up to 0.1 MHz)%

at 300 V subrange:
±0.01 % (up to 0.05 MHz) %
±0.02 % (up to 0.1 MHz) %

at 500 V subrange:
±0.02 % (up to 0.05 MHz)%
±0.03 % (up to 0.1 MHz)%

at 500 V subrange:
±0.02 % (up to 0.05 MHz)%
±0.03 % (up to 0.1 MHz)%

at 1000 V subrange:
±0.02 % (up to 0.05 MHz)%
±0.05 % (up to 0.1 MHz)%

at 300 V subrange:
±0.02 % (up to 0.05 MHz)%
±0.05 % (up to0.1 MHz)%

Input resistance and capacitance

at 0.5; 1; 3;
10; 30 V subranges:
> 200 Ohm/V; < 5 pF.
at 50; 100; 300; 500;
1000 V subranges:
> 1.2 kOhm/V; < 5 pF

at 0.5; 1; 3;
10; 30 V subranges:
> 200 Ohm/V; < 5 pF

at 50; 100; 300;
500; 1000 V subranges:
> 1.2 kOhm/V;