DESCRIPTION
The D1-25 Step Attenuator is a reference attenuator designed for testing and verification of measurement instrumentation in manufacturing environment.
The Instrument’s principle of operation is based on dividing dc and ac voltages by a precision resistor system, having low reactivity and high stability.
The D1-25 Step Attenuator provides 0 to 110 dB attenuation in 10-dB steps in 0–100 MHz operating frequency range.
The Instrument consists of a housing with attenuation sections and a switch with an indicator.
The coaxial input and output connectors are located in the housing, as well as additional «40 dB» and «80 dB» connectors for the Instrument’s verification.
N-type connector with cross-section of 7.0/3.04 mm, female, IEC STD 169-16.
SPECIFICATIONS
Frequency range: DC – 100 MHz
Dynamic attenuation range (in 10-dB steps): 0 to 110 dB
Basic differential attenuation error relative to zero mark at dc: ±(0.002 + 0.0002А) dB, where A is a differential attenuation
Permissible basic accuracy range in 0 to 30 MHz range:
±(0.004+a·f2) dB – for 10 dB differential attenuation relative to zero mark;
±(0.002+0.0002·А10+b·А10·f2) dB – for any differential attenuation value А10 dB, where a=0.0004 dB/MHz2, b=1·10–5 MHz–2, f – frequency in MHz.
Permissible basic accuracy of differential attenuation in 30 MHz to 100 MHz frequency range:
±(0.004+c·f), dB – for 10 dB differential attenuation relative to zero mark;
±(0.002+0.0002·А10+d·А10·f), dB – for any differential attenuation value А10, dB, relative to 10 dB mark, where c=0.012 dB/MHz, d=0.0003 MHz–1, f – frequency in MHz
DC input resistance: not more than 50±0.1 Ohm
DC output resistance (at any attenuation switch position into 50±0,1 Ohm load connected to the input): not more than 25±0.1 Ohm
Maximal dc or ac voltage applied to the input: not more than 3V
MTBF: not less than 30000 hours
Number of switching cycles from 0 to 110 dB: not less than 30 000
Dimensions: 149х149х95 mm.
GENERAL
Operating temperature range: 5–40 оС (non-operating temperature range –50 to 50 ºС)
Relative humidity at 25 оС: up to 98 %
Atmospheric pressure: 70 to 106.7 kPa